TECHNICAL DATA

Viscosity Range  

0.1 to 10,000 poise

Shear Rate Range

0.01 to 12000 sec -1

Cycle time

10 seconds upwards

Accuracy

better than 1%

Repeatability

better than 0.25%

Temperature Range

-20 to +300OC
(-4 to +572OF)

Maximum pressure at the capillary

100 bar (1500 psig) 

Maximum Vacuum which can be drawn from

-800 millibar (-22" mercury) relative to atmosphere 

Power Supply Requirements

110 Volts or 240 Volts 50 or 60 Hz
10 amps plus 6 amps for each additional head at 240 volts 

Air supply Requirements

Instrument air 5 - 10 bar
flow 2 litres/min. (max) per head. 

Communications

Intelligent programmable serial interface to give 

bulletRS232 
bulletRS485 
bulletTwo analogue outputs 
bullet0-20 mmA or 4-20mA at 0-5v or 1-5v 
bullet4 Digital outputs 
bullet(volt free contacts) 
bullet4 digital inputs (24v/48v) 
bulletPneumatic I/O options 0.2- 
bullet1 bar (3 - 15 psig) 

Maximum cable length between cubical and measuring head = 100 meters (110 yards) 

Dimensions

Measuring Head 600 x 250 x 200mm (w.d.h)
Control enclosure 600 x 600 x 1830 (w.d.h)
Weight : head 22Kg (50lbs)
Weight : Control enclosure 40Kg (270 lbs.) 

Options

Options on the materials of construction are available to give liquid compatibility, including non-taint materials for food products. Hazardous gas area specification is also available.